Operation theory: The Carbon Monoxide (CO) analyzer 9830 is a photometer in the non-dispersive infrared (NDIR), that uses the Gas Filters Correlation (GFC) in order to measure low concentrations of CO in a safely and reliable way.

The infrared radiation of wide range goes through a wheel of gas filter that contains half of CO and half of Nitrogen.

When the IR radiation passes through the half of the wheel containing CO, all the wave lengths absorbed by the CO are fully removed from the radiation, remaining the wave lengths that are not affected by the CO, giving a reference beam.

When the IR radiation passes through the half of the wheel containing Nitrogen, the wave lengths of specific absorption for CO are not removed from the radiation, giving a measurable beam, which will be dim by the presence of CO in the sample. The rotation of the wheel of gas filter produces a continuous alternative beam of reference and measure.

This batch beam (measure/reference) passes through a multipath absorption chamber (white chamber) where the measure of CO is made through the measure of the measure beam dim.
Due to the fact that both the reference and the measure beam come from the same transmitting source, have the same detector and cover the same optical path, the difference in measure of wave length between the reference and measure phases will only be affected by the CO quantity present in the sample. The measurement methodology gives as a result an instrument strongly insensitive to the interfering gases, fluctuations in the IR emission source, vibration and the tarnish of the optical parts.

The final concentration of CO corrected to changes of temperature and pressure is offered in ppm or milligrams per square meters (mg/m3).

Through the menus the user can configure the equipment and have access to all the options available.

The 9830 analyzer selects automatically the optimal range of measure both in the screen and in the RS-232 interface output. The values are indicated in the floating point mode, which makes unnecessary the selection and indication of the range of measure. The equipment includes a characteristic of analogical over range that, in activated mode, commutes the analogical output of the analyzer to a ranger higher than the previously defined.
 


Characteristics and functions

  • European manufacturer ISO9001 certified to ensure the supply of analyzers of quality and exceptional performance
  • The analyzers are approved and certified by the US-EPA and have been contrasted with reference standards by US, UK and TÜV and operate between +5ºC and +40ºC
  • The CO analyzer offers equivalent results as the obtained by reference methods commons in USA and CE (USA equivalent method as defined in 40 CFR Part 53)
  • The ML ® 9830B analyzer uses an IR source of long duration and a gas filters wheel manufactured in cutting edge materials to ensure a performance with no breakdowns and a minimum maintenance
  • The low flow of 1 l/min reduces the impact of the pollution of the fungibles and reduces the substitution costs and maintenance times
  • The equipment includes a self-diagnosis system and a full remote control through a RS-232 multipoint interface in order to optimize the service and maintenance availability
  • The analyzer has a wide LCD screen to watch simultaneously measure data and information regarding the working parameters
  • The analyzers from the 98xx series are noted for its modular design and for an internal performance at 12 Vcc with a consumption less than 100 W
  • The equipment has an internal register for relevant events as zero and calibration or specific error indicators up to 20 events
  • It incorporates internal memory in the main board. The data stored can be obtained in a remote mode
  • Kalman digital filter to give more exactitude and better response to real variations of the atmospheric pollutants concentrations compared with the filters of fixed wide beam of linear type

 


Specifications

  • Approval  Number: USA-EPA: RFCA-0992-088; TÜV: RW-TÜV3.5.1/554/92
  • Range:
    • Digital and screen output: Self-range from 0 to 200 ppm
  • Resolution = 0,01 ppm.
    • Analog output: 0 – full scale between 0-0,1 ppm and 0-200 ppm with 0, 5 and 10 % offset
    • Auto ranging analog output: two ranges selectable by the user: 0 – full scale
    • Range designed by USEPA: any range between 0-5 ppm 0-100 ppm
  • Noise:
    • Measure process: 0,025 ppm or 0,1% of the reading, whichever is greater (with Kalman algorithm active)
    • Analog interface: 0,025 ppm or 0,1 % of the analog full scale, whichever is greater
  • Lower Detectable Limit:
    • Measure process:<0,05 ppm or 0,2% of the reading, whichever the greater (with Kalman algorithm active)
    • Analog interfase:<0,05 ppm or 0,2% of the analog full scale, whichever the greater
  • Zero Drift:
    • Temperature dependent:<0,01 ppm/ ºC
    • Time dependent, at fixed temperature:
      • <0,1 ppm / 24 h
      • <0,1 ppm / 30 day
  • Span Drift:
    • Temperature dependent:<0,5 ppm/ ºC
    • Time dependent, at fixed temperature:
      • <0,5 %/24 hours of the concentration reading
      • <0,5 %/30 days of the concentration reading
  • Lag Time:<20 s
  • Rise/Fall Time T95:<40 sec (with Kalman algorithm active)
  • Error lineality: ±1% of the full scale
  • Precision: 0,1 ppm or 1 % of the concentration reading, whichever the greater
  • Sample Flow Rate: 1 l/min approx
  • Temperature Range:
    • TÜV: +5°C to +40°C (41°F to 104°F) -USEPA: +15°C to +35°C (59°F to 95°F)
  • Power Supply:
    • 99 to 132 VCA, 198 to 264 VCA, 47 to 63 Hz.
  • Weight: 21 Kg (46 lb)
  • Dimensions (w x h x d): 43,2 cm x 17,8 cm x 64,8 cm
  • Analog Output:
    • Menu selectable current output of 0 to 20 mA, 2 to 20 mA, or 4 to 20 mA
    • Optional voltage output with 50-pin connector board. Jumper selectable voltage output of 100 mV, 1 V, 5 V, and 10 V, with menu selectable zero offset of 0%, 5%, or 10%
  • Digital Output:
    • Multidrop RS232 port shared between analyzers for store of data, status and control