Low Volume Automatic Sampler CBV-30DSm/2,3. 30 filter capacity

 

 ACCORDING TO THE SPECIFICATIONS OF UNE-EN 12341:2015

Standardized reference method for PM10 and PM2,5

 

Low volume air sampler for determination of mass concentration  PM10 or PM2.5 of suspended particulate matter, filter sample for laboratory determination. Up to 30 filters.

Consists of a set of aspiration vane vacuum pump, suction head, filtration system and electronic system control assembled inside a polyester-fiberglass case outdoor suitable and  aluminum support frame that allows floor fixing.

Operation, measurement and flow regulation controlled by a microprocessor and a mass flow meter. Automatic control and constant maintenance of target flow rate with automatic compensation of the pressure drop for filter clogging or voltage fluctuations.

Flow measurment by orifice plate (according to EN 5167).

Continuous measurement of air temperature, humidity and atmospheric pressure. Sample temperature and atmospheric pressure. Sampled filter storage chamber temperature control. (Optionally it can incorporate other atmospheric sensors, like precipitation, speed and win direction and solar radiation).

Interface user via LCD graphic display 240 x 128 dots and keyboard. Internal memory for storage of data and accessible parameters by RS232 / USB, TCP/IP, SD-card and Wi-Fi. ModBus comunication Protocol admitted. Dot matrix printer included.

47mm Ø filter holder. Easy filter change by filter carrier holder. Filter holder valves closing by electromechanical system with filter position detector.

Forced ventilation and internal heating and cooling system for the sampled filter storage chamber under 20ºC with moisture removal and defrost.

PM10 or PM2,5 particles head built according to the specifications of UNE-EN 12341: 2015, with insulated tube connection and air sleeve.

 


Technical features

 

  • Flow: nominal of 2,3 m3/h. (real).
  • Regulation range: lower than 5% from nominal value.
  • Period flow average: deviation lower than 2% from nominal value.
  • Operating parameters visualization.
  • Sampling parameters and cycles programming.
  • Real time clock with an error of less than 1 minute/month.
  • Calibration parameters
  • Continuous measurement of air temperature, humidity and atmospheric pressure. Sample temperature and atmospheric pressure. Sampled filter storage chamber temperature control.
  • Assistance program for maintenance, fault detection and leak test.
  • Microprocessor electronic control.
  • Filter holder built in polyoxymethylene (POM) plastic and stainless steel filter support grid for 47 mm Ø filters.
  • PM10 or PM2,5 head. Anodized finish.
  • Assistance program for temperature, relative humidity and pressure sensor calibration
  • RS232/USB, TCP/IP, SD-card y WI-FI. ModBus comunication protocol admited.
  • Dot matrix printer included.
  • Power supply: 230V / 50 Hz. Energy consumption 350 VA max.
  • Sampler dimensions: 500 x 1040 x 420 mm (without head).
  • Support frame dimensions: 500 x 580 x 600 mm. 

Standardized head for PM10 and PM2,5 sampling

a)

    

 

b)

c)

a) PM10 and PM2,5 sampling head, connecting tube (inside) and air cover with forced ventilation. Pneumatic scheme:

 Sample airflow

 Air cover flow

b) PM10 Head

c) PM2,5 Head

 




© MCV, SA Terms Of Use